# Class of dictionnaries for Agilent IPCMini ion pump controller.
#
##########################################################################################
"""Contain the class of dictionnaries for Agilent IPCMini ion pump controller."""
import dataclasses
from types import MappingProxyType
[docs]@dataclasses.dataclass
class IonPumpDic:
"""Class of dictionnaries for Agilent IPCMini ion pump controller."""
dic_reply = MappingProxyType({6: "Command success", 15: "Command fail", 32: "Invalid window", 33: "Data type error", 34: "Out of range", 35: "Window is read-only"})
dic_mode = MappingProxyType({0: "Serial", 1: "Remote", 2: "Local", 3: "LAN"})
dic_baud_rate = MappingProxyType({1: 1200, 2: 2400, 3: 4800, 4: 9600})
dic_serial_type = MappingProxyType({False: "RS232", True: "RS485"})
dic_pressure_unit = MappingProxyType({0: "Torr", 1: "mBar", 2: "Pa"})
dic_device_number = MappingProxyType(
{
0: "Spare",
1: "500 StarCell",
2: "300 StarCell",
16: "200 StarCell",
3: "150 StarCell",
4: "75-55-40 StarCell",
5: "20 StarCell",
20: "NEXTorr-SC",
6: "500 diode",
7: "300 diode",
15: "200 diode",
8: "150 diode",
9: "75-55-40 diode",
10: "20 diode",
11: "10 diode",
12: "75 Sem",
13: "75 Sem",
14: "75 Sem",
17: "2 diode",
18: "0.2 diode 1250 Gauss",
19: "0.2 diode 800 Gauss",
},
)
dic_win = MappingProxyType(
{
"Mode": "008",
"HV ON/OFF": "011",
"Baud rate": "108",
"Status": "205",
"Error code": "206",
"Model": "319",
"Serial number": "323",
"RS485 address": "503",
"Serial type": "504",
"Unit pressure": "600",
"Autostart": "601",
"Protect": "602",
"Fixed/step": "603",
"Device number": "610",
"Max power": "612",
"V target": "613",
"I protect": "614",
"Set point": "615",
"Temperature power section": "800",
"Temperature internal controller": "801",
"Status set point": "804",
"V measured": "810",
"I measured": "811",
"Pressure": "812",
"Label": "890",
},
)
[docs]@dataclasses.dataclass
class IonPumpDics:
"""Class of joint dictionnaries for Agilent IPCMini ion pump controller."""
dics = MappingProxyType(
{
"Reply": IonPumpDic.dic_reply,
"Mode": IonPumpDic.dic_mode,
"Baud rate": IonPumpDic.dic_baud_rate,
"Serial type": IonPumpDic.dic_serial_type,
"Unit pressure": IonPumpDic.dic_pressure_unit,
"Device number": IonPumpDic.dic_device_number,
"Win": IonPumpDic.dic_win,
},
)
[docs]@dataclasses.dataclass
class IonPumpMinMaxDics:
"""Class of min/max values dictionnaries for Agilent IPCMini ion pump controller."""
dic_max_values = MappingProxyType(
{"LABEL_LENGTH": 10, "SETPOINT_CURRENT": 1.0e-5, "PROTECT_CURRENT": 10000, "VOLTAGE": 7000, "POWER": 40, "UNIT_PRESSURE": 2, "ADDRESS": 31, "BAUDRATE": 4, "MODE": 3},
)
dic_min_values = MappingProxyType({"SETPOINT_CURRENT": 1.0e-10, "PROTECT_CURRENT": 1, "VOLTAGE": 3000, "POWER": 10, "UNIT_PRESSURE": 0, "ADDRESS": 0, "BAUDRATE": 1, "MODE": 0})